FDA Enforcement Class II Terminated

RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions. The simulated plan is intended for treatment evaluation for example in tumor board meetings or operating rooms

Recall: Z-0647-2016 · Reported January 20, 2016

Enforcement

Recall Number
Z-0647-2016
Event ID
72804
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Brainlab AG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 20, 2016
Initiation Date
November 5, 2015
Classification Date
January 14, 2016
Termination Date
August 21, 2017
Address
Kapellenstr. 12, N/A, Feldkirchen, N/A, N/A, Germany

Description

RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions. The simulated plan is intended for treatment evaluation for example in tumor board meetings or operating rooms

Reason

Large objects with fine resolution are potentially displayed cropped when imported into Brainlab Brain Metastases 1.0.0.

Code Info

Brain Metastases version 1.0.0 Model/catalogue number Description 21702-04 ELEMENTS AUT. BRAIN METASTASES PLAN INST 21703-01 ELEMENTS ADAPT HYB SURG ANALYSIS LICENSE 21703-02 ELEMENTS ADAPT HYB SURG ANALYSIS INST 21702 ELEMENTS AUTOMATIC BRAIN METS PLAN SUB 21702-12 ELEMENTS AUTOMATIC BRAIN METS PLAN SUB 21702-36 ELEMENTS AUTOMATIC BRAIN METS PLAN SUB 21702-60 ELEMENTS AUTOMATIC BRAIN METS PLAN SUB 21703 ELEMENTS AD-HYBRID SURG ANALYSIS SUBSCR 21703-12 ELEMENTS AD-HYBRID SURG ANALYSIS SUBSCR 21703-36 ELEMENTS AD-HYBRID SURG ANALYSIS SUBSCR 21703-60 ELEMENTS AD-HYBRID SURG ANALYSIS SUBSCR 21713 ELEMENTS ADAPTIVE HYBRID SURG. ANALYSIS 21722 ELEMENTS AUTO. BRAIN MET PLAN SUBS. FOC 21722-12 ELEMENTS AUTO. BRAIN MET PLAN SUBS. FOC 21722-36 ELEMENTS AUTO. BRAIN MET PLAN SUBS. FOC 21722-60 ELEMENTS AUTO. BRAIN MET PLAN SUBS. FOC 21723 ELEMENTS ADAPT HYB SURG ANALYS SUBS. FOC 21723-12 ELEMENTS ADAPT HYB SURG ANALYS SUBS. FOC 21723-36 ELEMENTS ADAPT HYB SURG ANALYS SUBS. FOC 21723-60 ELEMENTS ADAPT HYB SURG ANALYS SUBS. FOC 21732 ELEMENTS AUTOMATIC BRAIN METS PLAN. 12M

Distribution

Worldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brazil, Canada, China, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Mexico, Portugal, South Korea, Spain, Switzerland, Turkey and the United Kingdom.

Quantity

20 systems