FDA Enforcement Class II Terminated

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Recall: Z-0641-2020 · Reported December 18, 2019

Enforcement

Recall Number
Z-0641-2020
Event ID
84216
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Integra LifeSciences Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 18, 2019
Initiation Date
October 21, 2019
Classification Date
December 8, 2019
Termination Date
May 7, 2020
Address
311 Enterprise Dr, N/A, Plainsboro, NJ, 08536-3344, United States

Description

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Reason

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Code Info

Finnish OPS Manual P/N 60905782

Distribution

US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK

Quantity

N/A