Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. UNIV.15DROP PUMP SET W/3 SAFELINE LL; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 362032. 2. UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; Catalog Number: 362033. 3. 60DROP METRISET PUMP SET, 3 SAFELINE; Priming volume: 19mL, length: 120in, SafeLine; Catalog Number: 362034. 4. Y TYPE BLOOD PUMP SET, SAFELINE INJ; Priming volume: 39mL, length: 120in, SafeLine¿; Catalog Number: 362043. 5. UNIV. 15 DROP PUMP SET W/ 0.2 FILTER; 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; Catalog Number: 362050. 6. SPACE PUMP SET 15 DROP W/2 SAFEDAY; SPACE PUMP SET 15 DROP W/2 SAFEDAY; Catalog Number: 362420. 7. UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL; 15 drops/mL, priming volume: 21mL, length: 120in, ULTRASITE; Catalog Number: 362431. 8. SPACE PUMP IV SET, 120 IN.; 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; Catalog Number: 362432. 9. 60 DROP METRISET PUMP SET,3 SFLINE ASV; Priming volume: 21mL, length: 121in, SafeLine; Catalog Number: 363010. 10. Y-TYPE BLOOD PUMP SET, SAFELINE, ASV; Priming volume: 40mL, length: 123in, SafeLine; Catalog Number: 363019. 11. UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 363030. 12. UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV; 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; Catalog Number: 363230. 13. Y-TYPE BLOOD PUMP SET, CARESITE, ASV; Priming volume: 41mL, length: 120in, CARESITE; Catalog Number: 363419. 14. SPACE PUMP IV SET, 2 CARESITES, ASV; 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; Catalog Number: 363420. 15. UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 363421. 16. 60DROP METRISET PUMP, 3 CARESITES, ASV; Priming volume: 21mL, length: 130in, CARESITE; Catalog Number: 363422. 17. UNIV. 60 DROP PUMP SET, 2 CARESITES, ASV; 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; Catalog Number: 363423. 18. UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; Catalog Number: 363424. 19. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; Catalog Number: 363430. 20. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; Catalog Number: 363433. 21. SPACE PUMP IV SET 15D, 110 IN.; "SPACE PUMP IV SET 15D, 110 IN. CLOSE"; Catalog Number: 480254. 22. SPACE PUMP IV SET, 127 IN.; SPACE PUMP IV SET, 127 IN.; Catalog Number: 480255. 23. SPACE PUMP SET 15D, CARESITE LAD 122 IN. ; 15 drops/mL, priming volume: 23mL, length: 122in, CARESITE; Catalog Number: 480256. 24. 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; Catalog Number: 480258. 25. SPACE PUMP IV SET 15D, 90 IN.; SPACE PUMP IV SET 15D, 90 IN.; Catalog Number: 480263. 26. UNIV. 15 DROP PUMP SET W 3 CARESITE L.L.; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 490100. 27. 60 DROP METRISET PUMP SET, 3 CARESITES; Priming volume: 19mL, length: 120in, CARESITE; Catalog Number: 490101.
Enforcement
- Recall Number
- Z-0640-2026
- Event ID
- 97816
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- B Braun Medical Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 3, 2025
- Initiation Date
- October 29, 2025
- Classification Date
- November 26, 2025
- Address
- 824 12th Ave, N/A, Bethlehem, PA, 18018-3524, United States
Description
Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. UNIV.15DROP PUMP SET W/3 SAFELINE LL; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 362032. 2. UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; Catalog Number: 362033. 3. 60DROP METRISET PUMP SET, 3 SAFELINE; Priming volume: 19mL, length: 120in, SafeLine; Catalog Number: 362034. 4. Y TYPE BLOOD PUMP SET, SAFELINE INJ; Priming volume: 39mL, length: 120in, SafeLine¿; Catalog Number: 362043. 5. UNIV. 15 DROP PUMP SET W/ 0.2 FILTER; 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; Catalog Number: 362050. 6. SPACE PUMP SET 15 DROP W/2 SAFEDAY; SPACE PUMP SET 15 DROP W/2 SAFEDAY; Catalog Number: 362420. 7. UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL; 15 drops/mL, priming volume: 21mL, length: 120in, ULTRASITE; Catalog Number: 362431. 8. SPACE PUMP IV SET, 120 IN.; 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; Catalog Number: 362432. 9. 60 DROP METRISET PUMP SET,3 SFLINE ASV; Priming volume: 21mL, length: 121in, SafeLine; Catalog Number: 363010. 10. Y-TYPE BLOOD PUMP SET, SAFELINE, ASV; Priming volume: 40mL, length: 123in, SafeLine; Catalog Number: 363019. 11. UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 363030. 12. UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV; 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; Catalog Number: 363230. 13. Y-TYPE BLOOD PUMP SET, CARESITE, ASV; Priming volume: 41mL, length: 120in, CARESITE; Catalog Number: 363419. 14. SPACE PUMP IV SET, 2 CARESITES, ASV; 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; Catalog Number: 363420. 15. UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 363421. 16. 60DROP METRISET PUMP, 3 CARESITES, ASV; Priming volume: 21mL, length: 130in, CARESITE; Catalog Number: 363422. 17. UNIV. 60 DROP PUMP SET, 2 CARESITES, ASV; 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; Catalog Number: 363423. 18. UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; Catalog Number: 363424. 19. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; Catalog Number: 363430. 20. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; Catalog Number: 363433. 21. SPACE PUMP IV SET 15D, 110 IN.; "SPACE PUMP IV SET 15D, 110 IN. CLOSE"; Catalog Number: 480254. 22. SPACE PUMP IV SET, 127 IN.; SPACE PUMP IV SET, 127 IN.; Catalog Number: 480255. 23. SPACE PUMP SET 15D, CARESITE LAD 122 IN. ; 15 drops/mL, priming volume: 23mL, length: 122in, CARESITE; Catalog Number: 480256. 24. 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; Catalog Number: 480258. 25. SPACE PUMP IV SET 15D, 90 IN.; SPACE PUMP IV SET 15D, 90 IN.; Catalog Number: 480263. 26. UNIV. 15 DROP PUMP SET W 3 CARESITE L.L.; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 490100. 27. 60 DROP METRISET PUMP SET, 3 CARESITES; Priming volume: 19mL, length: 120in, CARESITE; Catalog Number: 490101.
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
1. Catalog Number: 362032; Primary UDI-DI: 04046964182556; Unit of Dose UDI-DI: 04046964182549. Shelf Life: Earliest Exp of Comp or 60mths. 2. Catalog Number: 362033; Primary UDI-DI: 04046964837821; Unit of Dose UDI-DI: 04046964837814. Shelf Life: Earliest Exp of Comp or 36mths. 3. Catalog Number: 362034; Primary UDI-DI: 04046964182570; Unit of Dose UDI-DI: 04046964182563. Shelf Life: Earliest Exp of Comp or 48mths. 4. Catalog Number: 362043; Primary UDI-DI: 04046964182631; Unit of Dose UDI-DI: 04046964182624. Shelf Life: Earliest Exp of Comp or 60mths. 5. Catalog Number: 362050; Primary UDI-DI: 04046964182655; Unit of Dose UDI-DI: 04046964182648. Shelf Life: Earliest Exp of Comp or 60mths. 6. Catalog Number: 362420; Primary UDI-DI: 04046955592715; Unit of Dose UDI-DI: 04046955592708. Shelf Life: 36mths from Start Date. 7. Catalog Number: 362431; Primary UDI-DI: 04046964294013; Unit of Dose UDI-DI: 04046964294006. Shelf Life: Earliest Exp of Comp or 60mths. 8. Catalog Number: 362432; Primary UDI-DI: 04046964182679; Unit of Dose UDI-DI: 04046964182662. Shelf Life: Earliest Exp of Comp or 60mths. 9. Catalog Number: 363010; Primary UDI-DI: 04046964294174; Unit of Dose UDI-DI: 04046964294167. Shelf Life: Earliest Exp of Comp or 48mths. 10. Catalog Number: 363019; Primary UDI-DI: 04046964294198; Unit of Dose UDI-DI: 04046964294181. Shelf Life: Earliest Exp of Comp or 60mths. 11. Catalog Number: 363030; Primary UDI-DI: 04046964294259; Unit of Dose UDI-DI: 04046964294242. Shelf Life: Earliest Exp of Comp or 60mths. 12. Catalog Number: 363230; Primary UDI-DI: 04046964294495; Unit of Dose UDI-DI: 04046964294488. Shelf Life: Earliest Exp of Comp or 60mths. 13. Catalog Number: 363419; Primary UDI-DI: 04046964294570; Unit of Dose UDI-DI: 04046964294563. Shelf Life: Earliest Exp of Comp or 36mths. 14. Catalog Number: 363420; Primary UDI-DI: 04046964294594; Unit of Dose UDI-DI: 04046964294587. Shelf Life: Earliest Exp of Comp or 36mths. 15. Catalog Number: 363421; Primary UDI-DI: 04046964294617; Unit of Dose UDI-DI: 04046964294600. Shelf Life: Earliest Exp of Comp or 36mths. 16. Catalog Number: 363422; Primary UDI-DI: 04046964294631; Unit of Dose UDI-DI: 04046964294624. Shelf Life: Earliest Exp of Comp or 36mths. 17. Catalog Number: 363423; Primary UDI-DI: 04046964294655; Unit of Dose UDI-DI: 04046964294648. Shelf Life: Earliest Exp of Comp or 36mths. 18. Catalog Number: 363424; Primary UDI-DI: 04046964837906; Unit of Dose UDI-DI: 04046964837890. Shelf Life: Earliest Exp of Comp or 36mths. 19. Catalog Number: 363430; Primary UDI-DI: 04046964294679; Unit of Dose UDI-DI: 04046964294662. Shelf Life: 36mths from Start Date. 20. Catalog Number: 363433; Primary UDI-DI: 04046964837920; Unit of Dose UDI-DI: 04046964837913. Shelf Life: Earliest Exp of Comp or 36mths. 21. Catalog Number: 480254; Primary UDI-DI: 04046955324798; Unit of Dose UDI-DI: 04046955343904. Shelf Life: Earliest Exp of Comp or 36mths. 22. Catalog Number: 480255; Primary UDI-DI: 04046955324804; Unit of Dose UDI-DI: 04046955585403. Shelf Life: Earliest Exp of Comp or 36mths. 23. Catalog Number: 480256; Primary UDI-DI: 04046955324811; Unit of Dose UDI-DI: 04046955585861. Shelf Life: Earliest Exp of Comp or 36mths. 24. Catalog Number: 480258; Primary UDI-DI: 04046955324835; Unit of Dose UDI-DI: 04046955585885. Shelf Life: Earliest Exp of Comp or 36mths. 25. Catalog Number: 480263; Primary UDI-DI: 04046955693931; Unit of Dose UDI-DI: 04046955693924. Shelf Life: Earliest Exp of Comp or 36mths. 26. Catalog Number: 490100; Primary UDI-DI: 04046964186080; Unit of Dose UDI-DI: 04046964186073. Shelf Life: 36mths from Start Date. 27. Catalog Number: 490101; Primary UDI-DI: 04046964186103; Unit of Dose UDI-DI: 04046964186097. Shelf Life: Earliest Exp of Comp or 36mths. 28. Catalog Number: 490102; Primary UDI-DI: 04046964186127; Unit of Dose UDI-DI: 04046964186110. Shelf Life: Earliest Exp of Comp or 36mths. 29. Catalog Number: 490103; Primary UDI-DI: 04046964302107; Unit of Dose UDI-DI: 04046964302091. Shelf Life: Earliest Exp of Comp or 36mths. 30. Catalog Number: 490104; Primary UDI-DI: 04046964186141; Unit of Dose UDI-DI: 04046964186134. Shelf Life: Earliest Exp of Comp or 36mths. 31. Catalog Number: 490105; Primary UDI-DI: 04046964186165; Unit of Dose UDI-DI: 04046964186158. Shelf Life: Earliest Exp of Comp or 36mths. 32. Catalog Number: 490655; Primary UDI-DI: 04046955836383; Unit of Dose UDI-DI: 04046955836376. Shelf Life: Earliest Exp of Comp or 36mths.
Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
49,330,370