FDA Enforcement Class II Ongoing

Incisive CT, model 728143 & 728144 running Software Version 5.0.0.

Recall: Z-0640-2024 · Reported January 17, 2024

Enforcement

Recall Number
Z-0640-2024
Event ID
93574
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips Healthcare (Suzhou) Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 17, 2024
Initiation Date
December 23, 2022
Classification Date
January 5, 2024
Address
Suzhou Industrial Park, No. 258 Zhongyuan Road; Pingjiang, Suzhou, N/A, China

Description

Incisive CT, model 728143 & 728144 running Software Version 5.0.0.

Reason

Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan. These include: 1. Artifacts may occur on perfusion images, 2. Orientation label reversed in 2D images, 3. Auto voice does not work during the cardiac scan, 4. Auto voice and operators voice are too quiet, patient cant hear, 5. Not all of the images are reconstructed in some cases, 6. No image displayed on the console monitor after enabling CCT Single Scan, 7. Ring artifact appears in 32x0.625 collimation of bone scan when soft tissue view is selected, 8. A design defect where the error couch motion signal of couch firmware causes couch to identify current position as target position in next protocol, and 9. calcium scoring study split into separate series after loading to ISP.

Code Info

UDI-DI: 00884838085015 (Model 728143) UDI-DI: 00884838105508 (Model 728144)

Distribution

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, PA, RI, SC, TN, TX, UT, VT, WI, & Puerto Rico and the countries of Albania, Algeria, Argentina, Australia, Austria, Bhutan, Bolivia, Bosnia-Herz, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Mauritius, Moldova, Morocco, Netherlands, New Zealand, Norway, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Senegal, Serbia, Sint Maarten, Slovakia, South Africa, South Korea, Spain, St. Pier Miquel, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, United Arab Emirates, Uzbekistan, Venezuela, & Vietnam.

Quantity

828