FDA Enforcement Class II Terminated

VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifier Number 10758750009930, Software Product Code 6802866, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.

Recall: Z-0640-2016 · Reported January 20, 2016

Enforcement

Recall Number
Z-0640-2016
Event ID
72597
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 20, 2016
Initiation Date
November 3, 2015
Classification Date
January 13, 2016
Termination Date
May 30, 2018
Address
100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States

Description

VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifier Number 10758750009930, Software Product Code 6802866, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.

Reason

Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.

Code Info

Software Version 3.2 and below; Serial Numbers J36000012 - 36000879

Distribution

Worldwide Distribution - US Nationwide and the countries of Puerto Rico, and foreign distribution to: Argentina, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Israel, Italy, Japan, Kenya, Mexico, Panama, Philippines, Poland, Romania, Russia, Singapore, Spain, Taiwan, Turkey, and Venezuela.

Quantity

Domestic: 200, Foreign: 554