FDA Enforcement Class II Terminated

OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories

Recall: Z-0638-2022 · Reported February 23, 2022

Enforcement

Recall Number
Z-0638-2022
Event ID
89404
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Olympus Corporation of the Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 23, 2022
Initiation Date
December 13, 2021
Classification Date
February 11, 2022
Termination Date
December 26, 2023
Address
3500 Corporate Pkwy, N/A, Center Valley, PA, 18034-8229, United States

Description

OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories

Reason

The detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent for the cleaning cycle and detergent will be dispensed instead of alcohol for the alcohol cycle, resulting in a ineffective endoscope reprocessing and patient exposure to residual detergent

Code Info

All serial numbers. UDI: 04953170404047

Distribution

Nationwide

Quantity

295 units