FDA Enforcement Class II Terminated

Abbott m2000sp Instrument; an automated fluid handling system for performing sample preparation for nucleic acid testing; Abbott Molecular Inc., Des Plaines, IL 60018 USA; Abbott m2000sp Instrument; an automated system for performing sample preparation for nucleic acid testing; Abbott Molecular Inc., Des Plaines, IL 60018 USA; list 09K14 The Abbott m2000 System is intended for use in performing nucleic acid testing in clinical laboratories.

Recall: Z-0634-2013 · Reported January 9, 2013

Enforcement

Recall Number
Z-0634-2013
Event ID
63746
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Abbott Molecular
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 9, 2013
Initiation Date
December 4, 2012
Classification Date
January 3, 2013
Termination Date
March 8, 2016
Address
1300 E Touhy Ave, N/A, Des Plaines, IL, 60018-3315, United States

Description

Abbott m2000sp Instrument; an automated fluid handling system for performing sample preparation for nucleic acid testing; Abbott Molecular Inc., Des Plaines, IL 60018 USA; Abbott m2000sp Instrument; an automated system for performing sample preparation for nucleic acid testing; Abbott Molecular Inc., Des Plaines, IL 60018 USA; list 09K14 The Abbott m2000 System is intended for use in performing nucleic acid testing in clinical laboratories.

Reason

An issue with the m2000sp Instrument software has been identified whereby, after a series of steps, if changes are made to the control or calibrator information on the Sample Extraction Assay Details screen, the changes are not saved.

Code Info

list 09K14, all serial numbers and all software versions 1.0 - 5.0

Distribution

Worldwide Distribution - USA (nationwide) and internationally to Algeria, Argentina, Australia, Austria, Azerbaijan, Belgium, Bolivia, Brazil, Burkina Faso, Burundi, Cameroon, Canada, Chad, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ethiopia, Estonia, Finland, France, Gaza & Jericho, Germany, Greece, Guinea, Haiti, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Libya, Lithuania, Luxembourg, Malawi, Malaysia, Mali, Mexico, Mozambique, Netherlands, New Zealand, Norway, Pakistan, Panama, Paraguay, Peru, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam and Zambia

Quantity

215 units - U.S., 693 units - ROW