FDA Enforcement Class II Terminated

Perkin Elmer WIZARD2 1-detector, 550 samples Product Code: 2470-0010 The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples

Recall: Z-0632-2016 · Reported January 20, 2016

Enforcement

Recall Number
Z-0632-2016
Event ID
72808
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Perkinelmer
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 20, 2016
Initiation Date
December 8, 2015
Classification Date
January 12, 2016
Termination Date
August 4, 2017
Address
940 Winter Street, N/A, Waltham, MA, 02451, United States

Description

Perkin Elmer WIZARD2 1-detector, 550 samples Product Code: 2470-0010 The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples

Reason

The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.

Code Info

Serial Numbers: DG06106539 DG12118720 DG07106646 DG12107250 DG12107281 DG03117619 DG11118605 DG01117400 DG12107202 DG01117401 DG06106449 DG03117666 DG05117791 DG06106454 DG03106180 DG11118637 DG07118121 DG07106598 DG11118633 DG06106520 DG03106137 DG07118111 DG02128888 DG08118247 DG06118014 DG01128790 DG12107207 DG07106637 DG01128876 DG04117764 DG01106035 DG12118780 DG05106342 DG01117399 DG06117909 DG06118011 DG09106846 DG11107146 DG12118694 DG08118217 DG02106084 DG04117690 DG05117838 DG09106799 DG04117735 DG03117548 DG01117362

Distribution

Worldwide Distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.

Quantity

47 units