FDA Enforcement Class II Terminated

ABX PENTRA AST CP, Part Number: A11A01629. The ABX Pentra AST CP reagent is a diagnostic reagent for quantitative in vitro determination of Aspartate Amino Transferase (AST) in serum or plasma by colorimetry.

Recall: Z-0632-2013 · Reported January 9, 2013

Enforcement

Recall Number
Z-0632-2013
Event ID
63838
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Horiba Instruments, Inc dba Horiba Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 9, 2013
Initiation Date
October 10, 2012
Classification Date
January 3, 2013
Termination Date
September 12, 2013
Address
34 Bunsen, N/A, Irvine, CA, 92618-4210, United States

Description

ABX PENTRA AST CP, Part Number: A11A01629. The ABX Pentra AST CP reagent is a diagnostic reagent for quantitative in vitro determination of Aspartate Amino Transferase (AST) in serum or plasma by colorimetry.

Reason

Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.

Code Info

Lot number: 00456-4108

Distribution

Nationwide Distribution including CA, OK, NY, FL, AZ, NY, VA, and TX.

Quantity

66886 units