FDA Enforcement Class II Terminated

PKG, PENNINGTON FORCEPS, P/N 0250080242. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Recall: Z-0631-2015 · Reported December 24, 2014

Enforcement

Recall Number
Z-0631-2015
Event ID
69838
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Endoscopy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 24, 2014
Initiation Date
November 17, 2014
Classification Date
December 17, 2014
Termination Date
July 17, 2015
Address
5900 Optical Ct, N/A, San Jose, CA, 95138-1400, United States

Description

PKG, PENNINGTON FORCEPS, P/N 0250080242. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Code Info

345254 845624 846778 847452 944705 944722 1040097 1040577 1040975 1042679 1042686 1042826 1143022 1143976 1144207 1144323 1144660 1144945 1145225 1145388 1145817 1146346 1146676 1146687

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Quantity

180,573 units total