FDA Enforcement Class II Ongoing

BD Veritor Plus Analyzer - intended to provide rapid test results in near-patient settings. Catalog Number: 256066

Recall: Z-0630-2022 · Reported February 16, 2022

Enforcement

Recall Number
Z-0630-2022
Event ID
89358
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Becton Dickinson & Co.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 16, 2022
Initiation Date
January 5, 2022
Classification Date
February 9, 2022
Address
7 Loveton Cir, N/A, Sparks, MD, 21152-9212, United States

Description

BD Veritor Plus Analyzer - intended to provide rapid test results in near-patient settings. Catalog Number: 256066

Reason

May have the potential to overheat and/or cause fire, the issue can occur when an unauthorized alternative AC power adapter, and not the one provided by BD is used to power the instrument.

Code Info

All unexpired analyzers (Exp. Date 12/2021 - 8/2024) UDI: 00382902560661

Distribution

Worldwide distribution - US Nationwide and the countries of Armenia, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Fiji, Guatemala, Hong Kong, Honduras, Haiti, Indonesia, India, Jamaica, Cambodia, South Korea, Macao, Mexico, New Zealand, Panama, Peru, Philippines, Singapore, El Salvador, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Vietnam.

Quantity

126,952 Analyzers Updated:129, 848 Analyzers