FDA Enforcement Class II Terminated

Hemotherm¿ Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that supplies temperature controlled water to a heat exchanger. The device, in conjunction with a heat exchanger, is used to warm or cool the blood that enters and exits a blood heat exchanger for adult, pediatric or infant patients.

Recall: Z-0629-2018 · Reported February 21, 2018

Enforcement

Recall Number
Z-0629-2018
Event ID
78173
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cincinnati Sub-Zero Products LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 21, 2018
Initiation Date
November 27, 2017
Classification Date
February 15, 2018
Termination Date
July 2, 2018
Address
12011 Mosteller Rd, Cincinnati, OH, 45241-1528, United States

Description

Hemotherm¿ Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that supplies temperature controlled water to a heat exchanger. The device, in conjunction with a heat exchanger, is used to warm or cool the blood that enters and exits a blood heat exchanger for adult, pediatric or infant patients.

Reason

Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.

Code Info

none

Distribution

US Nationwide Distribution

Quantity