FDA Enforcement
Class III
Terminated
FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.
Recall: Z-0628-2016
·
Reported January 20, 2016
Enforcement
- Recall Number
- Z-0628-2016
- Event ID
- 72885
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Becton, Dickinson and Company, BD Biosciences
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 20, 2016
- Initiation Date
- December 14, 2015
- Classification Date
- January 11, 2016
- Termination Date
- October 31, 2016
- Address
- 2350 Qume Dr, N/A, San Jose, CA, 95131-1812, United States
Description
FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.
Reason
Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.
Code Info
Lot No. 4162520, 4206855, 4267919, 4276818, 4309977, 5009530, 5036624, 5085573, 5182818
Distribution
Nationwide Distribution
Quantity
1,388 units