FDA Enforcement
Class II
Terminated
Accu-Chek Connect Diabetes Management App, Catalog number 07562462001 / GTIN number 00365702700000 & Catalog number 07250452001 / GTIN number 00365702700017
Recall: Z-0625-2018
·
Reported February 21, 2018
Enforcement
- Recall Number
- Z-0625-2018
- Event ID
- 79001
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diabetes Care, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Other
- Report Date
- February 21, 2018
- Initiation Date
- June 8, 2017
- Classification Date
- February 15, 2018
- Termination Date
- July 27, 2018
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025, United States
Description
Accu-Chek Connect Diabetes Management App, Catalog number 07562462001 / GTIN number 00365702700000 & Catalog number 07250452001 / GTIN number 00365702700017
Reason
Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. Due to a software bug, when the OS region of the phone setting is changed, the unit of measure within the app may unexpectedly change. This creates a risk the app might not transfer the blood glucose result or the user might not correctly input numerical values for carbohydrate used for bolus advice.
Code Info
Software Versions: 1.2.0, 1.2.2, 1.2.3, 2.0.0, 2.0.1, and 2.1.0
Distribution
US Nationwide
Quantity
67,040 unique users of version 2.1.0