FDA Enforcement Class I Terminated

Arrow-Trerotola Over-The-Wire PTD Kit, Percutaneous Thrombolytic Device, REF PT-12709-WC, containing components: PT-12709-CA, PT-12709-C, The Arrow-Trerotola Percutaneous Thrombolytic Device (PTD) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venous (AV) fistulae and synthetic dialysis grafts.

Recall: Z-0623-2022 · Reported February 23, 2022

Enforcement

Recall Number
Z-0623-2022
Event ID
89390
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
TELEFLEX MEDICAL INC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 23, 2022
Initiation Date
January 11, 2022
Classification Date
February 15, 2022
Termination Date
September 11, 2024
Address
3015 Carrington Mill Blvd, N/A, Morrisville, NC, 27560-5437, United States

Description

Arrow-Trerotola Over-The-Wire PTD Kit, Percutaneous Thrombolytic Device, REF PT-12709-WC, containing components: PT-12709-CA, PT-12709-C, The Arrow-Trerotola Percutaneous Thrombolytic Device (PTD) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venous (AV) fistulae and synthetic dialysis grafts.

Reason

The firm received reports indicating PTD tip separation during use.

Code Info

Lot Number / UDI: 13F19K0564 / (01)20801902156758 (17)210930 (10)13F19K0564, Components: Code PT-12709-CA, Lot number 13C19K2024; Code PT-65709-HS, Lot number 14C19K1035; Code PT-03009, Lot number 13C19J1873; Lot Number / UDI: 13F20C0094 / (01)10801902121254 (17)220228 (10)13F20C0094, Components: Code PT-12709-C, Lot number 13C20C0498; Code PT-65709-HS, Lot number 14C19K1035; Code PT-03009, Lot number 13C19L0622; Code PT-03009, Lot number 13C19M0278; Code PT-03009, Lot number 13C19J1873; Lot Number / UDI: 13F20G0284 / (01)10801902121254 (17)220630 (10)13F20G0284, Components: Code PT-12709-C, Lot number 13C20G0230; Code PT-65709-HS, Lot number 14C20D1061; Code PT-03009, Lot number 13C20G0238; Code PT-03009, Lot number 13C20F2585; Lot Number / UDI: 13F20L0282 / (01)10801902121254 (17)221031 (10)13F20L0282, Components: Code PT-12709-C, Lot number 13C20J0307; Code PT-65709-HS, Lot number 14C20H0384; Code PT-03009, Lot number 13C20L0872; Lot Number / UDI 13F21A0497 / (01)10801902121254 (17)221231 (10)13F21A0497, Components: Code PT-12709-C, Lot number 13C20J0307; Code PT-65709-HS, Lot number 14C20J1453; Code PT-03009, Lot number 13C20M0144; Lot Number / UDI: 13F21A0718 / (01)10801902121254 (17)230131 (10)13F21A0718, Components: Code PT-12709-C, Lot number 13C21A1963; Code PT-65709-HS, Lot number 14C20J1453; Code PT-03009, Lot number 13C21A0855; Code PT-03009, Lot number 13C20M0144; Lot Number / UDI: 13X21E0008 / (01)10801902121254 (17)230630 (10)13F21F1187, Components: Code PT-12709-C, Lot number 13C21C1690; Code PT-65709-HS, Lot number 14C21D0635; Code PT-65709-HS, Lot number 14C21C0476; Code PT-03009, Lot number 13C21C2413; Lot Number / UDI: 13F21F1187 / (01)10801902121254 (17)230531 (10)13X21E0008, Components: Code PT-12709-C, Lot number 13X21E0009; Code PT-65709-HS, Lot number 14X21D0002; Code PT-03009, Lot number 13X21E0007

Distribution

Worldwide distribution: US Nationwide and the countries of AU, BE, CA, CL, CZ, DE, ES, FR, GB, IL, IT, KR, KW, NL, PA, PH, PT, SA, TR.

Quantity

676 units