FDA Enforcement Class II Completed

Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000

Recall: Z-0616-2024 · Reported January 3, 2024

Enforcement

Recall Number
Z-0616-2024
Event ID
93524
Classification
Class II
Status
Completed
Product Type
Devices
Firm
Landauer
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 3, 2024
Initiation Date
November 15, 2023
Classification Date
December 22, 2023
Address
2 Science Rd, Glenwood, IL, 60425-1531, United States

Description

Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000

Reason

MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.

Code Info

All serial numbers.

Distribution

US and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam

Quantity

584 units