Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
Recall: Z-0616-2024
·
Reported January 3, 2024
Enforcement
- Recall Number
- Z-0616-2024
- Event ID
- 93524
- Classification
- Class II
- Status
- Completed
- Product Type
- Devices
- Firm
- Landauer
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- January 3, 2024
- Initiation Date
- November 15, 2023
- Classification Date
- December 22, 2023
- Address
- 2 Science Rd, Glenwood, IL, 60425-1531, United States
Description
Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
Reason
MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.
Code Info
All serial numbers.
Distribution
US and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam
Quantity
584 units