FDA Enforcement Class II Terminated

Synthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement, cutting of bone and metal, drilling, decorticating, shaping, and smoothing of bones and teeth in a wide variety of surgical procedures, including, but not limited to general orthopaedic trauma, foot, hand, maxillofacial, neurosurgical, oral, otolaryngolical, reconstructive, and spine surgery.

Recall: Z-0615-2014 · Reported January 8, 2014

Enforcement

Recall Number
Z-0615-2014
Event ID
65007
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Synthes USA HQ, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 8, 2014
Initiation Date
April 15, 2013
Classification Date
December 31, 2013
Termination Date
September 14, 2016
Address
1302 Wrights Ln E, West Chester, PA, 19380-3417, United States

Description

Synthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement, cutting of bone and metal, drilling, decorticating, shaping, and smoothing of bones and teeth in a wide variety of surgical procedures, including, but not limited to general orthopaedic trauma, foot, hand, maxillofacial, neurosurgical, oral, otolaryngolical, reconstructive, and spine surgery.

Reason

Synthes received a complaint from a customer in which the Hand Switch for the Electric Pen Drive was not labeled correctly. The "on" and "lock" labels were interchanged.

Code Info

Part Number 05.001.012, lot number AV18166

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to Mexico and Canada. .

Quantity

290