FDA Enforcement
Class II
Terminated
COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland
Recall: Z-0614-2013
·
Reported January 2, 2013
Enforcement
- Recall Number
- Z-0614-2013
- Event ID
- 63814
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Molecular Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 2, 2013
- Initiation Date
- October 11, 2012
- Classification Date
- December 22, 2012
- Termination Date
- September 11, 2013
- Address
- 1080 Us Highway 202 S, N/A, Branchburg, NJ, 08876-3733, United States
Description
COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland
Reason
Certain COBAS AmpliPrep instruments using AMPILINK software v.3.3.4 had the sample clot detection flagging feature disabled during quality control release resulting in samples that may not be flagged as having clots.
Code Info
CAP 394436 to 394663
Distribution
Distributed in the states of GA, CA, NC, and MN.
Quantity
7 units