FDA Enforcement Class II Terminated

Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Usage: is a device for measuring the refractive power of contact lenses and spectacle lenses and displaying measurement values for sphere, cylinder and prism.

Recall: Z-0613-2021 · Reported December 23, 2020

Enforcement

Recall Number
Z-0613-2021
Event ID
86366
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Carl Zeiss Meditec AG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 23, 2020
Initiation Date
August 6, 2020
Classification Date
December 16, 2020
Termination Date
December 13, 2024
Address
Carl-Zeiss-Promenade 10, N/A, Jena, N/A, N/A, Germany

Description

Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Usage: is a device for measuring the refractive power of contact lenses and spectacle lenses and displaying measurement values for sphere, cylinder and prism.

Reason

Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

Code Info

Model: VISUREF 150 Catalog Number: 000000-2227-967 Serial Numbers: Pending

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia, Canada, Colombia, Croatia, Cyprus, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Latvia, Luxembourg, Malaysia, Maldives, Netherlands, Portugal, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Turkey, and United Kingdom.

Quantity

603 units