FDA Enforcement
Class III
Terminated
Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.
Recall: Z-0608-2013
·
Reported January 2, 2013
Enforcement
- Recall Number
- Z-0608-2013
- Event ID
- 37530
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Radiometer America Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 2, 2013
- Initiation Date
- June 6, 2006
- Classification Date
- December 21, 2012
- Termination Date
- February 5, 2013
- Address
- 810 Sharon Dr, N/A, Westlake, OH, 44145-1598, United States
Description
Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.
Reason
The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.
Code Info
Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.
Distribution
Distributed to KY, NC, and NY.
Quantity
840 Ampules