FDA Enforcement Class III Terminated

Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.

Recall: Z-0608-2013 · Reported January 2, 2013

Enforcement

Recall Number
Z-0608-2013
Event ID
37530
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Radiometer America Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 2, 2013
Initiation Date
June 6, 2006
Classification Date
December 21, 2012
Termination Date
February 5, 2013
Address
810 Sharon Dr, N/A, Westlake, OH, 44145-1598, United States

Description

Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.

Reason

The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

Code Info

Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.

Distribution

Distributed to KY, NC, and NY.

Quantity

840 Ampules