Flexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL1-HC, Order Number G49045 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.
Enforcement
- Recall Number
- Z-0607-2021
- Event ID
- 86804
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cook Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 30, 2020
- Initiation Date
- November 24, 2020
- Classification Date
- December 23, 2020
- Termination Date
- September 7, 2022
- Address
- 750 N Daniels Way, N/A, Bloomington, IN, 47404-9120, United States
Description
Flexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL1-HC, Order Number G49045 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
Lot Numbers: 13105055, 13124400, 13124401, 13135880, 13138926, 13140911, 13141335, 13162741, 13162742, 13168155, 13168156, 13170665, 13189283, 13214450, 13214451, 13214452, 13214455, 13221812, 13231810, 13237657, 13249434, 13252354, 13252362, 13262332, 13288118, 13291868, 13294511, 13307182, 13315176, 13337784, 13337785, 13337786, 13346090, 13354241, 13357216, 13360416, 13366208, 13366240, 13380385, 13388332
US Nationwide Distribution including states of: AK, CT, IA, LA, MO, NH, OK, TN, WI, AL, DC, ID, MA, MS, NJ, OR, TX, WV, AR, DE, IL, MD, MT, NM, PA, UT, WY, AZ, FL, IN, ME, NC, NV, RI, VA, CA, GA, KS, MI, ND, NY, SC, VT, CO, HI, KY, MN, NE, OH, SD, WA and Global Distribution to: AU, SG, Austria, Belgium, Canary Islands, CN, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Ireland, Italy, KR, Latvia, Lithuania, MY, Morocco, Netherlands, Norway, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, TW, TH, and United Arabs Emirates.
1527