Lyophilized Chemistry Calibrator Level 1 Product Usage: Beckman Coulter Chemistry Calibrators are intended for use when calibrating methods run on the Beckman Coulter AU series of chemistry analyzers. Beckman Coulter Chemistry Calibrators are lyophilized, human serum based products formulated for use as a reference material when calibrating Beckman Coulter AU clinical chemistry system assays. The calibrators are used to calibrate the following assays: Albumin (ALB), Bicarbonate (CO2), Direct Bilirubin (DBIL), Total Bilirubin (TBIL), Calcium Arsenazo (CALA), Calcium (CAL oCPC), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Phosphorus (PHOS), Iron, Lactate, Magnesium (MG), Total Protein (TP), Triglyceride (TRIG), Unbound Iron Binding Capacity (UIBC), Urea Nitrogen (BUN), and Uric Acid. Only lots 6101K51 and 6101K61 were subject to recall.
Enforcement
- Recall Number
- Z-0605-2018
- Event ID
- 78699
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 21, 2018
- Initiation Date
- August 24, 2017
- Classification Date
- February 13, 2018
- Termination Date
- August 28, 2020
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232, United States
Description
Lyophilized Chemistry Calibrator Level 1 Product Usage: Beckman Coulter Chemistry Calibrators are intended for use when calibrating methods run on the Beckman Coulter AU series of chemistry analyzers. Beckman Coulter Chemistry Calibrators are lyophilized, human serum based products formulated for use as a reference material when calibrating Beckman Coulter AU clinical chemistry system assays. The calibrators are used to calibrate the following assays: Albumin (ALB), Bicarbonate (CO2), Direct Bilirubin (DBIL), Total Bilirubin (TBIL), Calcium Arsenazo (CALA), Calcium (CAL oCPC), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Phosphorus (PHOS), Iron, Lactate, Magnesium (MG), Total Protein (TP), Triglyceride (TRIG), Unbound Iron Binding Capacity (UIBC), Urea Nitrogen (BUN), and Uric Acid. Only lots 6101K51 and 6101K61 were subject to recall.
Unpaired lots of AU lyophilized chemistry calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results. QC failures could occur if incorrect pairs of calibrators are used.
Lots # 6101K51 and 6101K61
Worldwide Distribution in the states of AL, AK, AZ, AK, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IO, KS, KY, LA, MA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NY, NM, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VI, WA, WV, WI, and WY and the countries of Brazil, Canada, Costa Rica, El Salvador, Guam, Paraguay, and Peru
unknown