FDA Enforcement Class II Terminated

Proximal Humerus Plate, 8 holes, Right, REF MDS130108R - Product Usage: The Proximal Humerus Plating System consists of plates, screws, and locking caps intended to provide internal fixation of fractures of the proximal humerus. The polyaxial screws have a range of motion of +/- 20¿. The screws can be simply lagged or locked to the plate with the addition of a locking cap. The plates are side specific implants.

Recall: Z-0603-2020 · Reported December 11, 2019

Enforcement

Recall Number
Z-0603-2020
Event ID
84230
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Miami Device Solutions, LLC.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
December 11, 2019
Initiation Date
January 15, 2019
Classification Date
December 5, 2019
Termination Date
December 11, 2020
Address
7620 Nw 25th St Ste 3, Miami, FL, 33122-1719, United States

Description

Proximal Humerus Plate, 8 holes, Right, REF MDS130108R - Product Usage: The Proximal Humerus Plating System consists of plates, screws, and locking caps intended to provide internal fixation of fractures of the proximal humerus. The polyaxial screws have a range of motion of +/- 20¿. The screws can be simply lagged or locked to the plate with the addition of a locking cap. The plates are side specific implants.

Reason

Right Proximal Humerus Plates. Product component parts MDS130108R, may be improperly identified as left.

Code Info

Lot number PM00208

Distribution

US Nationwide distribution in the states of FL, RI, MA, PA, NY, OH, GA, WA, AL, CA, TX, VA.

Quantity

15 units