FDA Enforcement Class II Ongoing

GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Artist system, nuclear magnetic resonance imaging system.

Recall: Z-0600-2022 · Reported February 16, 2022

Enforcement

Recall Number
Z-0600-2022
Event ID
89376
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
GE Healthcare, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 16, 2022
Initiation Date
December 23, 2021
Classification Date
February 9, 2022
Address
3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States

Description

GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Artist system, nuclear magnetic resonance imaging system.

Reason

The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.

Code Info

All systems.

Distribution

Distribution was nationwide, including Puerto Rico and the Virgin Islands. There was also government and military distribution. Foreign distribution was made to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaidjan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Republic of Congo, Croatia, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Equatorial Guinea, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrhyzstan, Latvia, Lebanon, Libyan Arabic Jamahiriya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Mali, Malta, Mauritania, Mexico, Republic of Moldova, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Nordic-HQ, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Zambia, and Zimbabwe.

Quantity

17,228 devices total in this field correction