FDA Enforcement Class II Terminated

UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotract

Recall: Z-0598-2019 · Reported December 19, 2018

Enforcement

Recall Number
Z-0598-2019
Event ID
81576
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Neotract Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 19, 2018
Initiation Date
July 6, 2018
Classification Date
December 12, 2018
Termination Date
December 2, 2020
Address
4473 Willow Rd Ste 100, N/A, Pleasanton, CA, 94588-8570, United States

Description

UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotract

Reason

Upon implant deployment, the Capsular Tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a capsular tab may result in a delay in completing a treatment or an inability to complete a treatment for the patient. Use of a recalled device may lead to increased frequency of exiting known risks associated with the product, including bleeding and tissue trauma associated with delivering the needle.

Code Info

Model: AN00141 (US) and AN00142 (OUS); Catalog No.: UL400-4; Lots: P37940, P37943, P37953, P37956, P37965, P37971, P37985, P37988, P37995, P38001, P38013, P38020, P38124, P38132

Distribution

US: AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV,

Quantity

8,804