FDA Enforcement Class II Terminated

Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.

Recall: Z-0597-2016 · Reported January 13, 2016

Enforcement

Recall Number
Z-0597-2016
Event ID
72751
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 13, 2016
Initiation Date
November 18, 2015
Classification Date
January 6, 2016
Termination Date
December 5, 2016
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.

Reason

Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.

Code Info

Software Version: SW VD-10E

Distribution

Nationwide Distribution including: AL, AZ, CA, CO, FL, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NC, NY, OH, OK, PA, SC, TN, TX, VA, and WI.

Quantity

73 units