FDA Enforcement
Class II
Terminated
Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.
Recall: Z-0597-2016
·
Reported January 13, 2016
Enforcement
- Recall Number
- Z-0597-2016
- Event ID
- 72751
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 13, 2016
- Initiation Date
- November 18, 2015
- Classification Date
- January 6, 2016
- Termination Date
- December 5, 2016
- Address
- 40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States
Description
Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.
Reason
Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.
Code Info
Software Version: SW VD-10E
Distribution
Nationwide Distribution including: AL, AZ, CA, CO, FL, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NC, NY, OH, OK, PA, SC, TN, TX, VA, and WI.
Quantity
73 units