FDA Enforcement Class I Ongoing

Catalog No. 1443 DRESSING CHANGE TRAY / STERILE - LATEX FREE

Recall: Z-0593-2024 · Reported January 10, 2024

Enforcement

Recall Number
Z-0593-2024
Event ID
93546
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Busse Hospital Disposables, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
January 10, 2024
Initiation Date
November 30, 2023
Classification Date
January 3, 2024
Address
75 Arkay Dr, N/A, Hauppauge, NY, 11788-3707, United States

Description

Catalog No. 1443 DRESSING CHANGE TRAY / STERILE - LATEX FREE

Reason

Firm is recalling all kits containing recalled lots of Nurse Assist Sterile 0.9% Normal Saline. Nurse Assist identified product in distribution for which the required sterility assurance level (SAL) of 10^-6 cannot be guaranteed. Risks to patients if product is used include blood stream infections, urinary tract infections, or respiratory infections; in the worst-case scenario, there is risk of sepsis or death.

Code Info

UDI-DI 00849233010727 Lots 2210077 2210112 2310275 2310331

Distribution

US Nationwide distribution.

Quantity

536