FDA Enforcement
Class II
Terminated
InSure ONE Test Strips (25 pack) vials-fecal immunochemical test (FIT) qualitatively detects human hemoglobin from blood in fecal samples Model 12045.01.
Recall: Z-0590-2021
·
Reported December 23, 2020
Enforcement
- Recall Number
- Z-0590-2021
- Event ID
- 86674
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Enterix, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 23, 2020
- Initiation Date
- October 8, 2020
- Classification Date
- December 15, 2020
- Termination Date
- January 10, 2023
- Address
- 236 Fernwood Ave, N/A, Edison, NJ, 08837-3839, United States
Description
InSure ONE Test Strips (25 pack) vials-fecal immunochemical test (FIT) qualitatively detects human hemoglobin from blood in fecal samples Model 12045.01.
Reason
Product does not meet internal testing criterion; use of the product may result in a potential false positive test result and may lead to patients undergoing unnecessary follow-up diagnostic procedure, i.e., colonoscopy or sigmoidoscopy
Code Info
Lot Numbers: 770G11 and 770G21
Distribution
Worldwide distribution - US Nationwide and the country of Australia.
Quantity
US: 5,256 vials (131,400 test strips); OUS:1051 vials