RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, including all service packs. RayStation 10A and RayPlan 10A are also affected, but not the 10A service pack. (Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0)
Enforcement
- Recall Number
- Z-0589-2021
- Event ID
- 86832
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- RAYSEARCH LABORATORIES AB
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 23, 2020
- Initiation Date
- November 11, 2020
- Classification Date
- December 15, 2020
- Termination Date
- May 9, 2023
- Address
- Sveavagen 9, N/A, Stockholm, N/A, N/A, Sweden
Description
RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, including all service packs. RayStation 10A and RayPlan 10A are also affected, but not the 10A service pack. (Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0)
Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan 9B and 10A, 6D couch angles may be incorrect.
Software version numbers: 4.0.0.14 , 4.0.3.4 , 4.5.1.14 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 4.7.6.7 , 4.9.0.42 ,; 5.0.1.11 , 5.0.2.35 , 5.0.3.17 , 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154
US Nationwide distribution.
124