FDA Enforcement
Class II
Terminated
DIST LAT FEM LOCK PLT, RT
Recall: Z-0588-2014
·
Reported January 8, 2014
Enforcement
- Recall Number
- Z-0588-2014
- Event ID
- 66499
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- January 8, 2014
- Initiation Date
- November 5, 2013
- Classification Date
- December 27, 2013
- Termination Date
- May 5, 2015
- Address
- 345 E Main St, Warsaw, IN, 46580-2746, United States
Description
DIST LAT FEM LOCK PLT, RT
Reason
Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure
Code Info
Item Numbers: 47-2357-101-14 47-2357-101-18 47-2357-102-14 47-2357-102-18 All lots with an expiration date prior to July 24, 2023
Distribution
Worldwide Distribution.
Quantity
42, 065 units total