FDA Enforcement Class II Terminated

DIST LAT FEM LOCK PLT, LT

Recall: Z-0587-2014 · Reported January 8, 2014

Enforcement

Recall Number
Z-0587-2014
Event ID
66499
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 8, 2014
Initiation Date
November 5, 2013
Classification Date
December 27, 2013
Termination Date
May 5, 2015
Address
345 E Main St, Warsaw, IN, 46580-2746, United States

Description

DIST LAT FEM LOCK PLT, LT

Reason

Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure

Code Info

Item Number: 47-2357-102-14 47-2357-102-18 All lots with an expiration date prior to July 24, 2023

Distribution

Worldwide Distribution.

Quantity

42, 065 units total