ADULT Radiotransparent Electrode, Part number C100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.
Enforcement
- Recall Number
- Z-0583-2015
- Event ID
- 69875
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Heart Sync, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 31, 2014
- Initiation Date
- November 11, 2014
- Classification Date
- December 23, 2014
- Termination Date
- November 24, 2015
- Address
- 5643 Plymouth Rd, Ann Arbor, MI, 48105-9586, United States
Description
ADULT Radiotransparent Electrode, Part number C100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.
Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the
All Lots. Catalog / Part Number C100-PHILIPS
Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and South Korea.
sold in boxes of ten electrodes: 4663.8 boxes; 46638 units total