FDA Enforcement Class II Ongoing

Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.

Recall: Z-0582-2025 · Reported December 11, 2024

Enforcement

Recall Number
Z-0582-2025
Event ID
95755
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Wilson-Cook Medical Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 11, 2024
Initiation Date
October 31, 2024
Classification Date
December 2, 2024
Address
4900 Bethania Station Rd, Winston Salem, NC, 27105-1203, United States

Description

Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.

Reason

Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.

Code Info

UDI-DI 00827002246638, Lot numbers W4844245, exp. 6/10/2027; W4847501, exp. 6/20/2027; and W4851589, exp. 7/8/2027.

Distribution

Distribution was OUS only to Argentina, Brazil, and Israel.

Quantity

46 devices