FDA Enforcement Class II Terminated

Remote Patient Monitoring System, MDP 2000 Series, Part Numbers: MDP2040-0100, PGM340R3, or PGM340R4. The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected Examples of areas where the MDP200 is used include ICU, CCU, PACU, Emergency, Telemetry Step-down and other areas where patient's physiological information are to be observed at one or more locations. Typical information displayed includes ECG and blood pressure waveforms and numeric values which as heart rate and systolic, mean and diastolic pressure. This device acts as the Central Monitor Station, it permits from one (1) to sixteen (16) bedside monitor(s) to be connected simultaneously.

Recall: Z-0582-2016 · Reported January 13, 2016

Enforcement

Recall Number
Z-0582-2016
Event ID
72794
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Vidco, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
January 13, 2016
Initiation Date
November 8, 2015
Classification Date
January 4, 2016
Termination Date
August 16, 2016
Address
6175 Sw 112th Ave, N/A, Beaverton, OR, 97008-4838, United States

Description

Remote Patient Monitoring System, MDP 2000 Series, Part Numbers: MDP2040-0100, PGM340R3, or PGM340R4. The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected Examples of areas where the MDP200 is used include ICU, CCU, PACU, Emergency, Telemetry Step-down and other areas where patient's physiological information are to be observed at one or more locations. Typical information displayed includes ECG and blood pressure waveforms and numeric values which as heart rate and systolic, mean and diastolic pressure. This device acts as the Central Monitor Station, it permits from one (1) to sixteen (16) bedside monitor(s) to be connected simultaneously.

Reason

Testing at customer site showed unit Remote Patient Monitoring System MDP2040-0100 in a continuous trap condition, not allowing system to reset and reboot. Two customers recently complained of the system freezing and it could only be restarted if the user re-applied power.

Code Info

Serial numbers: 6533-6537, 6561, 6566-6581, 6539, 6540, 6558-6560, 6562, 6508-6528, 6563-6565, 6529, 6531, 6532, 6541, 6551, 6556, and 6557.

Distribution

US Nationwide distribution in the states of AZ, CA, MD, NM, NJ, and OH.

Quantity

59 units