FDA Enforcement Class II Terminated

REFLECTION(R), MICROSTABLE(R), ACETABULAR LINER, 20o OVERHANG, 28 MM ID, 54-56 OD, QTY: (1), STERILE EO, REF 71742854, Smith & Nephew, Inc. Memphis, TN 38116 USA. Orthopaedic.

Recall: Z-0581-2014 · Reported January 1, 2014

Enforcement

Recall Number
Z-0581-2014
Event ID
67042
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Smith & Nephew Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 1, 2014
Initiation Date
November 25, 2013
Classification Date
December 26, 2013
Termination Date
February 1, 2017
Address
1450 E. Brooks Rd, N/A, Memphis, TN, 38116, United States

Description

REFLECTION(R), MICROSTABLE(R), ACETABULAR LINER, 20o OVERHANG, 28 MM ID, 54-56 OD, QTY: (1), STERILE EO, REF 71742854, Smith & Nephew, Inc. Memphis, TN 38116 USA. Orthopaedic.

Reason

One batch of the Reflection(TM) Polyethylene Acetabular Liners may be out of tolerance causing the liner not to fully seat into the shell.

Code Info

Batch No. 12BM04836

Distribution

Distributed in China and Spain.

Quantity

38 units