FDA Enforcement Class II Ongoing

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Recall: Z-0580-2025 · Reported December 4, 2024

Enforcement

Recall Number
Z-0580-2025
Event ID
95624
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Micro-X Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 4, 2024
Initiation Date
October 22, 2024
Classification Date
November 27, 2024
Address
Unit 14 6 Mab Eastern Promenade, Tonsley, N/A, Australia

Description

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Reason

Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.

Code Info

UDI-DI: 9357123000037, Serial Numbers: 348, 349, 361, 351, 365, 391

Distribution

US: TX, PR, MD, CA

Quantity

6