FDA Enforcement Class II Terminated

EMS products from Graham Medical, SnugFit(R) Expandable Mattress Cover, Code 72930, UPC 1 07 47036 72930 3

Recall: Z-0578-2018 · Reported February 21, 2018

Enforcement

Recall Number
Z-0578-2018
Event ID
78899
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
LITTLE RAPIDS CORPORATION
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 21, 2018
Initiation Date
December 27, 2017
Classification Date
February 9, 2018
Termination Date
July 30, 2018
Address
2273 Larsen Rd, Green Bay, WI, 54303-4809, United States

Description

EMS products from Graham Medical, SnugFit(R) Expandable Mattress Cover, Code 72930, UPC 1 07 47036 72930 3

Reason

The secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient transfer, not lifting. However, we recognize that the product may be utilized outside of its intended design scope, and lifting may occur. Lifting could lead tearing of the equipment cover and the potential for patient falling through.

Code Info

Lot codes: 17871, 17971, 18071, and 18971

Distribution

TX, CT, TN, MO, WI, FL, SD, MS, CA, MS

Quantity

153 cases, 30 units in each case