FDA Enforcement Class II Terminated

ARCHITECT HAVAB-G Calibrator, REF 6L27-01

Recall: Z-0574-2020 · Reported December 4, 2019

Enforcement

Recall Number
Z-0574-2020
Event ID
84238
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Abbott Gmbh & Co. KG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 4, 2019
Initiation Date
November 7, 2019
Classification Date
November 25, 2019
Termination Date
August 31, 2023
Address
Max-Planck-Ring 2, 65205, Wiesbaden, N/A, Germany

Description

ARCHITECT HAVAB-G Calibrator, REF 6L27-01

Reason

Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient results. Patient values generated using the impacted calibrator lot 08219BE00 may have changed interpretation from reactive to falsely non reactive. Results impacted may present as a (falsely) non reactive result that fall in the range between 0.50 to 0.99 S/CO.

Code Info

lot number 08219BE00

Distribution

US in the states of: AZ, CA, CO, CT, FL, KY, LA, MA, MD, MI, MN, MO, NJ, NY, OH, OR, RI, SC, TN, TX, WA, WV

Quantity

83 kits