ARCHITECT HAVAB-G Calibrator, REF 6L27-01
Recall: Z-0574-2020
·
Reported December 4, 2019
Enforcement
- Recall Number
- Z-0574-2020
- Event ID
- 84238
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott Gmbh & Co. KG
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 4, 2019
- Initiation Date
- November 7, 2019
- Classification Date
- November 25, 2019
- Termination Date
- August 31, 2023
- Address
- Max-Planck-Ring 2, 65205, Wiesbaden, N/A, Germany
Description
ARCHITECT HAVAB-G Calibrator, REF 6L27-01
Reason
Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient results. Patient values generated using the impacted calibrator lot 08219BE00 may have changed interpretation from reactive to falsely non reactive. Results impacted may present as a (falsely) non reactive result that fall in the range between 0.50 to 0.99 S/CO.
Code Info
lot number 08219BE00
Distribution
US in the states of: AZ, CA, CO, CT, FL, KY, LA, MA, MD, MI, MN, MO, NJ, NY, OH, OR, RI, SC, TN, TX, WA, WV
Quantity
83 kits