Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate nephrostomy balloon catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned on the catheter inside the balloon that define the working length
Enforcement
- Recall Number
- Z-0572-2026
- Event ID
- 97757
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Olympus Corporation of the Americas
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 26, 2025
- Initiation Date
- October 17, 2025
- Classification Date
- November 20, 2025
- Address
- 3500 Corporate Pkwy, Center Valley, PA, 18034-8229, United States
Description
Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate nephrostomy balloon catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned on the catheter inside the balloon that define the working length
Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit.
Model: BPCN0815K; UDI: 00821925033016; Lot Number: 25037006;
US Domestic: AK, CA, IA, KY, WA, WI; OUS International: Canada and Germany.
60 units