FDA Enforcement Class II Ongoing

Injection Needle, 3 Fr., REF: 27030N; Injection Cannula, 6 Fr., REF: 27182A

Recall: Z-0572-2024 · Reported December 27, 2023

Enforcement

Recall Number
Z-0572-2024
Event ID
93328
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Karl Storz Endoscopy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 27, 2023
Initiation Date
October 24, 2023
Classification Date
December 18, 2023
Address
2151 E Grand Ave, N/A, El Segundo, CA, 90245-5017, United States

Description

Injection Needle, 3 Fr., REF: 27030N; Injection Cannula, 6 Fr., REF: 27182A

Reason

The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.

Code Info

REF/UDI-DI: 27030N/4048551257180, 27182A/4048551081457 All lot numbers.

Distribution

US: MI, NH, SC, PA, FL, NY, NC, MS, MO, CA, NJ, IL, NV, OK, CO, OH, WA, WI, VA, IN, IA, AR, TX, WV, MD, MN, GA, MA, NM, VT, AK, OR, ID, LA, KS, DC, KY, HI OUS: Mexica, Canada, Germany

Quantity

990