FDA Enforcement Class II Ongoing

Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software

Recall: Z-0571-2025 · Reported December 4, 2024

Enforcement

Recall Number
Z-0571-2025
Event ID
95665
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
LivaNova Deutschland GmbH
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 4, 2024
Initiation Date
October 18, 2024
Classification Date
November 27, 2024
Address
Lindberghstr. 25, Munich, N/A, Germany

Description

Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software

Reason

Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.

Code Info

UDI-DI: 04033817903062. Software Version: 1.5. Serial Numbers: 49PK00436, 49PK00108, 49PK00374, 49PK00479, 49PK00480, 49PK00481, 49PK00093, 49PK00091, 49PK00092, 49PK00163, 49PK00497, 49PK00498, 49PK00219, 49PK00223, 49PK00224, 49PK00225, 49PK00226, 49PK00241, 49PK00255, 49PK00292, 49PK00291, 49PK00293, 49PK00490, 49PK00487, 49PK00488, 49PK00278, 49PK00274, 49PK00277, 49PK00275, 49PK00276, 49PK00273, 49PK00457, 49PK00458, 49PK00459, 49PK00455, 49PK00456, 49PK00290, 49PK00288

Distribution

US Nationwide distribution in the states of CA, VA, GA, IN, NJ, FL, CO, MS, NV.

Quantity

38