FDA Enforcement Class II Terminated

Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The HeartStart MRx EtCO2 Option monitors carbon dioxide (CO2) and measures end-tidal carbon dioxide (EtCO2). The option provides an EtCO2 value and a CO2 waveform on the devices display.

Recall: Z-0570-2014 · Reported January 1, 2014

Enforcement

Recall Number
Z-0570-2014
Event ID
67046
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 1, 2014
Initiation Date
December 10, 2013
Classification Date
December 24, 2013
Termination Date
April 18, 2017
Address
3000 Minuteman Rd, N/A, Andover, MA, 01810-1032, United States

Description

Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The HeartStart MRx EtCO2 Option monitors carbon dioxide (CO2) and measures end-tidal carbon dioxide (EtCO2). The option provides an EtCO2 value and a CO2 waveform on the devices display.

Reason

MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg

Code Info

Software versions: F.02, R.02, 9.xx or any earlier versions Addendum Models: M3532A, M3535A, M3536AM, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 M3535A- HeartStart MRx Defibrillator Monitor M3536A- HeartStart MRx Defibrillator Monitor EtC02 Options A04 A05 A06 All M3532A Upgrade

Distribution

US Nationwide and OUS.

Quantity

18,529 units