FDA Enforcement
Class II
Ongoing
Microsurgical Operating Room Tables;7300BX (18-351-70).
Recall: Z-0567-2021
·
Reported December 23, 2020
Enforcement
- Recall Number
- Z-0567-2021
- Event ID
- 85442
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Mizuho America, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- December 23, 2020
- Initiation Date
- September 11, 2019
- Classification Date
- December 14, 2020
- Address
- 30057 Ahern Ave, N/A, Union City, CA, 94587-1234, United States
Description
Microsurgical Operating Room Tables;7300BX (18-351-70).
Reason
Unintended tilting of the Tabletop MST 7300BX and incorrect operation due to infrared operating signal errors of Tabletops MST 7300B and MST 7300BX.
Code Info
Tilt Movement-Leaking Valve (7300BX-18-351-70) serial numbers: MUC-20163-0001, MUC-2016Z-0002, MUC-2016Z-0003, MUC-20177-0005, MUC-20177-0007, MUC-20177-0008, MUC-20177-0009, MUC-20189-0017 & MUC-2018Y-0018. Unintended movement due to infrared signal errors (7300BX-18-351-70) serial numbers: MUC-20163-0001, MUC-2016Z-0002, MUC-2016Z-0003, MUC-20177-0005, MUC-20177-0007, MUC-20177-0008, MUC-20177-0009, MUC-20189-0017, MUC-2018Y-0018 & MUC-2018Y-0019.
Distribution
US: AZ, CA, FL, MO, NC, OH, PA, TX, and WI. OUS: Greece, Israel, Japan, and Ukraine.
Quantity
10