FDA Enforcement Class II Terminated

Medtronic ANGLED DRILL, REF 6630902, QTY: 1 EA Rx only

Recall: Z-0566-2018 · Reported February 21, 2018

Enforcement

Recall Number
Z-0566-2018
Event ID
78839
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medtronic Sofamor Danek USA Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 21, 2018
Initiation Date
December 22, 2017
Classification Date
February 9, 2018
Termination Date
July 31, 2018
Address
1800 Pyramid Pl, Memphis, TN, 38132-1703, United States

Description

Medtronic ANGLED DRILL, REF 6630902, QTY: 1 EA Rx only

Reason

Medtronic has discovered the product lot in question was noted as having a longer drill tip extension than what is indicated on the print specification.

Code Info

UDI: 00643169363502, Lot Number EM17F015

Distribution

MN, NC

Quantity

2 units