FDA Enforcement Class II Terminated

Transpac¿ IV Disposable Transducer w/Stopcocks w/o Cap and Covers, Item No. 46112-30NS The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a compatible pressure monitor. The SafeSet Reservoir system is a blood sampling system assembled directly to the patient monitoring line (Transpac units). The SafeSet reservoir is used on patients where periodic blood sampling is required.

Recall: Z-0563-2017 · Reported November 23, 2016

Enforcement

Recall Number
Z-0563-2017
Event ID
75491
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 23, 2016
Initiation Date
September 27, 2016
Classification Date
November 15, 2016
Termination Date
July 11, 2017
Address
951 Calle Amanecer, N/A, San Clemente, CA, 92673-6212, United States

Description

Transpac¿ IV Disposable Transducer w/Stopcocks w/o Cap and Covers, Item No. 46112-30NS The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a compatible pressure monitor. The SafeSet Reservoir system is a blood sampling system assembled directly to the patient monitoring line (Transpac units). The SafeSet reservoir is used on patients where periodic blood sampling is required.

Reason

ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.

Code Info

3265805 3281660

Distribution

Nationwide Distribution

Quantity

180 units