FDA Enforcement Class I Ongoing

Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A, Sterile; (5) Medline Mini Tracheostomy Care Kit, REF CC3T3030A, Sterile; and (6) Medline Mini Tracheostomy Care Kit, REF DYND40622, Non-sterile.

Recall: Z-0561-2024 · Reported January 3, 2024

Enforcement

Recall Number
Z-0561-2024
Event ID
93327
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
MEDLINE INDUSTRIES, LP - Northfield
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 3, 2024
Initiation Date
October 16, 2023
Classification Date
December 22, 2023
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A, Sterile; (5) Medline Mini Tracheostomy Care Kit, REF CC3T3030A, Sterile; and (6) Medline Mini Tracheostomy Care Kit, REF DYND40622, Non-sterile.

Reason

The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.

Code Info

(1) DYND40580 - UDI 10080196032761, lot numbers: 6052203006, 6052203022, 6052204005, 6052204010, 6052205024, 6052206003, 6052206010, 6052206011, 6052206020, 6052206021, 6052207009, 6052207013, 6052207014, 6052207015,6052208021, 6052208022, 6052209009, 6052209025, 6052209026, 6052212023, 6052301031, 6052303011, 6052303013, 6052303014, 6052304020, 6052305007, 6052305019, 6052306004, 6052306014, 6052306015, 6052305026, 6052306026, 6052306032, 6052307006, 6052307011, and 6052307012. (2) DYND4061030 - UDI 10080196032877, lot numbers: 6052209025, 6052209026, 6052210016, 6052210031, 6052210032, 6052211034, 6052212009, 6052212010, 6052212021, 6052212022, 6052212015, 6052301017, 6052306014, 6052306015, 6052307011, 6052307012, 6052307022, 6052307023, 6052307026, 6052308001, 6052308003, 6052309001, 6052309003, and 6052309002. (3) CC3T4691A - UDI 10080196503391, lot numbers: 6052302012, 6052302013, 6052306004, 6052306014, 6052306015, 6052306017, 6052307012, 6052307022, 6052307023, 6052308002, and 6052308003. (4) CC4681A - UDI 10080196503407, lot numbers: 6052301017, 6052302012, 6052302013, 6052302028, 6052302029, 6052304024, 6052305020, 6052306004, 6052306014, 6052306015, 6052307011, and 6052307012. (5) CC3T3030A - UDI 10080196503384, lot numbers: 6052204010, 6052204011, 6052205021, 6052205022, 6052205023, 6052205024, 6052205028, 6052206003, 6052206010, 6052206011, 6052210015, 6052210016, 6052210031, 6052210032, 6052212010, 6052212023, 6052301015, 6052301016, 6052301017, 6052301029, 6052302012, 6052302013, 6052303011, 6052303012, 6052303013, 6052303014, 6052304019, 6052304020, 6052305007, 6052305020, 6052306004, 6052306014, 6052306015, 6052307011, 6052307012, 6052307022, 6052307023, 6052308001, 6052308002, and 6052308003. (6) DYND40622 - UDI 10080196032914, lot numbers: 6052203006, 6052204005, 6052204010, 6052205023, 6052205028, 6052206003, 6052206010, 6052207009, 6052207030, 6052208009, 6052208021, 6052208040, 6052209027, 6052209025, 6052210008, 6052210031, 6052210032, 6052212009, 6052212021, 6052212023, 6052301016, 6052302013, 6052302028, 6052303012, 6052303014, 6052305018, 6052305019, 6052305020, 6052306004, 6052306014, 6052306015, 6052306017, and 6052307022.

Distribution

US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.

Quantity

620,735 kits