MicroScan Pos Combo Panel Type 34, Part No. B1017-214 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.
Enforcement
- Recall Number
- Z-0560-2016
- Event ID
- 72779
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 6, 2016
- Initiation Date
- December 1, 2015
- Classification Date
- December 30, 2015
- Termination Date
- July 29, 2016
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
MicroScan Pos Combo Panel Type 34, Part No. B1017-214 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.
Beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) with Quality Control (QC) American Type Culture Collection (ATCC) organism Staphylococcus aureus ATCC 29213 in a portion of affected lots of MicroScan Pos Combo and MicroScan Pos Breakpoint Combo panels. The expected result is positive.
2016-04-29 2016-05-06 Part No. B1017-214
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Austria, Belgium, Brazil, Canada, Germany, Italy, Japan, Kazakhstan, Kuwait, Mexico, Myanmar, Netherlands, Norway, Panama, Poland, Portugal, Romania, Russian Federation, Spain, Taiwan, Thailand and United Kingdom.
1,324 units total (1,215 units in US)