FDA Enforcement Class II Terminated

MicroScan Pos Combo Panel Type 33, Part No. B1017-211 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.

Recall: Z-0559-2016 · Reported January 6, 2016

Enforcement

Recall Number
Z-0559-2016
Event ID
72779
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 6, 2016
Initiation Date
December 1, 2015
Classification Date
December 30, 2015
Termination Date
July 29, 2016
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

MicroScan Pos Combo Panel Type 33, Part No. B1017-211 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.

Reason

Beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) with Quality Control (QC) American Type Culture Collection (ATCC) organism Staphylococcus aureus ATCC 29213 in a portion of affected lots of MicroScan Pos Combo and MicroScan Pos Breakpoint Combo panels. The expected result is positive.

Code Info

2016-04-29 2016-05-04 Part No. B1017-211

Distribution

Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Austria, Belgium, Brazil, Canada, Germany, Italy, Japan, Kazakhstan, Kuwait, Mexico, Myanmar, Netherlands, Norway, Panama, Poland, Portugal, Romania, Russian Federation, Spain, Taiwan, Thailand and United Kingdom.

Quantity

2,706 units total (1,276 units in US)