FDA Enforcement Class II Terminated

Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Recall: Z-0558-2019 · Reported December 12, 2018

Enforcement

Recall Number
Z-0558-2019
Event ID
81545
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 12, 2018
Initiation Date
October 26, 2018
Classification Date
December 5, 2018
Termination Date
May 12, 2020
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reason

This recall has been initiated due to a possible leakage of a hose in the cooling unit of the X-ray tube. The Artis zee system may lose coolant due to a possible leakage in a hose of the cooling unit in the X-ray tube. This will cause the X-ray tube not to be sufficiently cooled down. As a result, the system will display a message Tube hot, have a break . The X-ray will be blocked several minutes later and to prevent any damage to the machine the following message will be displayed NO XRAY, TUBE TOO HOT . This issue is very sporadic and may occur during an ongoing clinical procedure. Use of this product may have a potential impact on planned procedures as they may be delayed as well as in emergent cases, as clinical treatment may need to be terminated and transferred to a functioning system.

Code Info

Serial nos: 910045-09 911354-02 910045-19 908883-23 910045-15 910045-14 910045-25 908883-04 911354-25 908883-18 908883-11 910045-21 911354-26 908883-06 908883-02 911354-27

Distribution

US Distribution to states of: AR, CA, KS, KY, MD, MN, NM, OK, PA, RI, TX, and UT.

Quantity

64 (16 in US)