FDA Enforcement Class II Ongoing

Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with Injection Port 8 mm x 40 mm; Cardiovascular. Product Numbers CH08-40-75GB and CH08-40-75US

Recall: Z-0557-2022 · Reported February 9, 2022

Enforcement

Recall Number
Z-0557-2022
Event ID
89418
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Covidien Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 9, 2022
Initiation Date
December 17, 2021
Classification Date
February 3, 2022
Address
15 Hampshire St, N/A, Mansfield, MA, 02048-1113, United States

Description

Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with Injection Port 8 mm x 40 mm; Cardiovascular. Product Numbers CH08-40-75GB and CH08-40-75US

Reason

The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloon is damaged while removing the protector sleeve and the catheter is subsequently used in a patient procedure, patient harm may potentially occur, including perforation of vessels, foreign body in patient, embolism, risks associated with surgical intervention, or delay to treatment/therapy.

Code Info

UDI: 10884521812253, 07290016745047, 10884521812291 For CFN CH08-40-75US: GTIN Batch 07290016745047 21904374 07290016745047 21904432 07290016745047 22000673 07290016745047 22000709 07290016745047 22000817 07290016745047 22000880 07290016745047 22001030 07290016745047 22001061 07290016745047 22001191 07290016745047 22001310 07290016745047 22001417 07290016745047 22001448 07290016745047 22001728 07290016745047 22002086 07290016745047 22002088 10884521812253 22002498 07290016745047 22002583 10884521812253 22002603 07290016745047 22003037 07290016745047 22100003 07290016745047 22100321 07290016745047 22100322 07290016745047 22101089 07290016745047 22101090 10884521812291 22102189

Distribution

Worldwide distribution - US Nationwide and the countries of Germany, Israel, Italy, Portugal, Spain, Chile, Singapore, and China.

Quantity